Abstract:Tomographic imaging systems are expected to work with a wide range of samples that house complex structures and challenging material compositions, which can influence image quality in a bad way. Complex samples increase total measurement duration and may introduce beam-hardening artifacts that lead to poor reconstruction image quality. This work presents an online trajectory optimization method for an X-ray computed tomography system with a robotic sample holder. The proposed method reduces measurement time and increases reconstruction image quality by generating an optimized spherical trajectory for the given sample without prior knowledge. The trajectory is generated successively at runtime based on intermediate sample measurements. We present experimental results with the robotic sample holder where two sample measurements using an optimized spherical trajectory achieve improved reconstruction quality compared to a conventional spherical trajectory. Our results demonstrate the ability of our system to increase reconstruction image quality and avoid artifacts at runtime when no prior information about the sample is provided.
Abstract:This work presents methods for the seamless execution of arbitrary spherical trajectories with a seven-degree-of-freedom robotic arm as a sample holder. The sample holder is integrated into an existing X-ray computed tomography setup. We optimized the path planning and robot control algorithms for the seamless execution of spherical trajectories. A precision-manufactured sample holder part is attached to the robotic arm for the calibration procedure. Different designs of this part are tested and compared to each other for optimal coverage of trajectories and reconstruction image quality. We present experimental results with the robotic sample holder where a sample measurement on a spherical trajectory achieves improved reconstruction quality compared to a conventional circular trajectory. Our results demonstrate the superiority of the discussed system as it outperforms single-axis systems by reaching nearly 82\% of all possible rotations. The proposed system is a step towards higher image reconstruction quality in flexible X-ray CT systems. It will enable reduced scan times and radiation dose exposure with task-specific trajectories in the future, as it can capture information from various sample angles.