Computerized Tomography assembly and system configuration are optimized for enhanced invertibility in sparse data reconstruction. Assembly generating maximum principal components/condition number of weight matrix is designated as best configuration. The gamma CT system is used for testing. The unoptimized sample location placement with 7.7% variation results in a maximum 50% root mean square error, 16.5% loss of similarity index, and 40% scattering noise in the reconstructed image relative to the optimized sample location when the proposed criteria are used. The method can help to automate the CT assembly, resulting in relatively artifact-free recovery and reducing the iteration to figure out the best scanning configuration for a given sample size, thus saving time, dosage, and operational cost.